The BC Biomedical ESU-2050P adds a Pulse Mode to the ESU-2050.
The calibration quality analyzer uses external precision non-inductive test loads and wide band current transformer. The standard model is compatible with all legacy generators with continuous output waveforms, and with the latest technology generators such as the Covidien Force Triad©, Ligasure Blood Vessel Dealing generator, etc. The standard model is NOT compatible with generators that produce pulsed output waveforms.
This device requires a Pearson Coil and External Resistors (not included).
• A CQM-2000 is helpful if you do not have another means of testing CQM/REM/ARM.
• 1% measurement accuracy
• Uses external current transformer & loads
• mV, mV Peak, mA, crest factor & wattage displays
• Waveform capture, storage, display & download
• Graphical display with bright white backlight
• Price includes: Universal Power Supply, Soft Carrying Case, RS-232 & USB Communication Cables, BNC to BNC Cable (3-foot) for Current
• Transformer connection, PC utility software